Simrad ER60 - REV B Operator's Manual page 37

Scientific echo sounder application
Table of Contents

Advertisement

Figure 4 Target
strength of a 60 mm
copper sphere
850-164692 / Rev.B
Copper spheres
Frequency
Diameter
(kHz)
(mm)
12*
45.0
18
63.0
27
42.0
38
60.0
49*
45.0
50*
45.0
70
32.1
120
23.0
200
13.7
710
10.3
(*) = same sphere
The curve for the 38 kHz sphere is given as an illustration.
1400
1420
60 mm copper sphere, 38 kHz
Calibration using ER60
Only one frequency may be calibrated at a time. Thus, you must
uninstall all transceivers in ER60 except for the one you intend
to calibrate. When you have calibrated this frequency you must
uninstall the transceiver, install the next transceiver, and
calibrate this frequency, and so on until all frequencies are
calibrated.
All the different combinations of pulse duration and transmitter
power that will be used during normal operation of the echo
sounder must be calibrated.
First, create a well organised file directory for the calibration
results. Example:
c:\er60\calibration\2003\38kHz\0.256msec\
Operational procedures
TS
(sound speed = 1490 m/s) (dB)
--40.4
--34.4
--37.9
--33.6
--36.4
--36.2
--39.1
--40.4
--45.0
--50.5
Pulse duration (ms)
1440
1460
1480
1500
SOUND SPEED (m/s)
3.0
0.3
1520
1540
1560
29

Advertisement

Table of Contents
loading

This manual is also suitable for:

Er60

Table of Contents