30
The same structure can be repeated for the different pulse
durations for each frequency. When you save the calibration file,
create a file name that includes the current date and the current
output power. Save the information as a TXT file. Example:
....\0.256msec\cal010603-2000w.txt
While the calibration takes place, we recommend that you also
collect a raw data file. Browse the file directory in the Store
dialogue box to locate the correct directory. The raw data file
name will be generated automatically, but you may specify a
prefix. The raw data file is used for replay.
The same procedure can be repeated for each of the installed
frequencies.
Observe the following calibration procedure.
1
Ensure that only one transceiver is installed in the ER60
echo sounder software.
2
From the Operation menu, select Ping control.
3
In the Ping Control dialogue, set Ping rate to Interval
and 1 second.
- This can also be done from the Toolbar.
4
From the Operation menu, select Normal.
5
In the Normal Operation dialogue, select transceiver, and
switch to Active mode.
6
In the Normal Operation dialogue, set the Transmit
Power to the level you wish to calibrate.
7
In the Normal Operation dialogue, choose the Pulse
Duration you wish to calibrate.
8
Right-click in an echogram, select Range on the shortcut
menu, and set the range for one of the echogram views to
cover the range you wish to see.
- This range should include the depth range where you
expect to find the reference target.
9
Check that you see the reference target in the Single Echo
view.
10
Click the right mouse button in the Single Echo view
corresponding to the echogram to open the Single Target
Detection dialogue box.
11
In the Single Target Detection dialogue, click the
Calibration button to start the calibration program, and to
create a new Calibration window.
- The calibration program allows you to record new
calibration data, or read previously recorded calibration
data.