6.3 Troubleshooting and Fault Diagnosis

Co d e
Description
M9
Simulated maintenance fault
M10
Over Range
M11
User gas calibration expired.
M13
Flow error.
M14
RESERVED
M15
Temperature near limit.
M16
Baseline fault.
M17
Inhibit timeout.
M93
M94
LIT pump error
F39
Simulated fault
F40
RESERVED
F41
Baseline fault.
F42
Calibration expired.
F43
Cartridge expired.
F44
Cell failure.
F45
RESERVED
F46
Cartridge analog failure.
F47
Cartridge memory invalid.
Midas S2
The user has generated a simulated fault.
A large concentration has been detected.
The Midas S2 requires independent
confirmation that the gas hazard is gone.
The user-specified calibration
Midas S2 can no longer regulate flow
(high or low flow condition).
The temperature is within
2 Celsius of the limits.
The sensor baseline has drifted.
The transmitter has been in
inhibit mode for too long.
LIT fail
Leakage on the sample line during LIT
calibration or LIT test
The pressure sensor does not work
The user has generated a simulated fault.
The sensor baseline has drifted.
Too long since the last calibration.
The cartridge needs to be updated.
Cartridge failed the Reflex™ check.- ECC.
Cartridge failed the CAT check.- FL/IR
Condition
interval has elapsed.
-
properly
-
-
Electronic Failure.
Checksum error.
46
Recovery
Reset simulated fault.
Supply known clean air to the Midas S2 and clear this
fault.
Perform zero and span calibrations. Increase span calibration
period.
Check filters and pump.
Check pressure at the inlet and outlet and
assure they are within specification.
-
Check the installation environment.
Check for background gas concentration,
temperature, or humidity fluctuations.
Perform zero calibration. Replace cartridge.
Resume monitoring or increase the timeout value.
Secure sample line leakage with a blocking test and check the
connection of the sampling line. If there is no leakage, service or
replace the main module.
Service or replace the main module.
Reset simulated fault.
-
Check for background gas concentration, temperature,
or humidity fluctuations. Replace cartridge.
Replace or calibrate the cartridge.
Replace cartridge.
Replace cartridge.
-
Replace cartridge.
Replace cartridge.