Finishing; Troubleshooting; Waste Electrical And Electronic Equipment (Weee) - Honeywell Midas S2 User Manual

Fixed single-point extractive gas transmitter
Hide thumbs Also See for Midas S2:
Table of Contents

Advertisement

6.1.3 Finishing

Wait for the gas reading to return to normal levels. This may take some time, especially if
concentrations above the recommended levels have been used or if the test gas has been
applied for an extended period of time.
Ensure that any latched faults or alarms that have been generated by the testing have been
cleared.
Make sure that the gas detection system is fully restored to its normal operating state and that
any control system overrides used during the testing have been removed.
The alarms generated by the testing will have created entries in the transmitter's history log.

6.1.4 Troubleshooting

If the test does not produce a satisfactory result, check the following points:
Check the gas cylinder concentration and type are correct
l
Check the expiration date of the cylinder.
l
Check that there is sufficient gas left in the cylinder.
l
Check that there are no leaks on the test system.
l
Make sure that the tubing length is as short as possible and that there are no blockages.
l
Make sure that the transmitter is not displaying a fault before the test is started.
l

6.2 Waste Electrical and Electronic Equipment (WEEE)

EU Directive 2012/19/EU
Midas S2
This symbol indicates that the product must not be disposed of as general industrial or domestic waste. This
product should be disposed of through suitable WEEE disposal facilities. For more information about
disposal of this product, contact your local authority, distributor or the manufacturer.
45
User Manual