Ec / Tds Probe Maintenance - Hanna Instruments HI 23 Series Instruction Manual

Wall-mounted, microprocessor-based, conductivity and tds process controllers
Hide thumbs Also See for HI 23 Series:
Table of Contents

Advertisement

EC / TDS PROBE MAINTENANCE

PERIODIC MAINTENANCE
CLEANING PROCEDURE
Note
Probe can be compensated for normal contamination by a
process of recalibration. When calibration can no longer
be achieved, remove the conductivity probe from the sys-
tem for maintenance.
Inspect the probe and the cable. The cable used for the
connection to the controller must be intact and there must
be no points of broken insulation on the cable.
Connectors must be perfectly clean and dry.
Rinse the probe with tap water. If a more thorough cleaning
is desired, remove the sleeve and clean the platinum sen-
sors with a nonabrasive cloth or HI7061 cleaning solution.
Reinsert the sleeve in the same direction as before.
Recalibrate the instrument before reinserting the probe in
the system.
Always recalibrate the instrument when attaching a new
probe.
55