EC VALUES AT VARIOUS TEMPERATURES

54
Temperature has a significant effect on conductivity. Table
below shows EC values at various temperatures for the
Hanna calibration solutions.
TEMPERATURE
EC VALUES (µS/cm)
°C
°F
HI7030
HI7031 HI7033 HI7034 HI7035
HI8030
HI8031 HI8033 HI8034 HI8035
0
32
7150
776
5
41
8220
896
10
50
9330
1020
15
59
10480
1147
16 60.8
10720
1173
17 62.6
10950
1199
18 64.4
11190
1225
19 66.2
11430
1251
20
68
11670
1278
21 69.8
11910
1305
22 71.6
12150
1332
23 73.4
12390
1359
24 75.2
12640
1386
25
77
12880
1413
26 78.8
13130
1440
27 80.6
13370
1467
28 82.4
13620
1494
29 84.2
13870
1521
30
86
14120
1548
31 87.8
14370
1575
HI7039
HI8039
64
48300
65400
2760
65
53500
74100
3180
67
59600
83200
3615
68
65400
92500
4063
70
67200
94400
4155
71
68500
96300
4245
73
69800
98200
4337
74
71300
100200
4429
76
72400
102100
4523
78
74000
104000
4617
79
75200
105900
4711
81
76500
107900
4805
82
78300
109800
4902
84
80000
111800
5000
86
81300
113800
5096
87
83000
115700
5190
89
84900
117700
5286
90
86300
119700
5383
92
88200
121800
5479
94
90000
123900
5575

Advertisement