5.1 - General

Temperature Range:
OS137A-1
OS137A-2
OS137A-3
Accuracy:
OS137A-1, OS137-2
OS137A-3
Repeatability:
Field of View:
Spectral Response:
Response Time:
Emissivity:
Alarm Output:
Alarm Set Point:
Analog output:
MA
V1
V2
K
MVC
MVF
Output Load Requirements:
Min. Load
Min. Load
Max. Load
Min. Load
Min. Load
Operating Ambient Temperature:
No Water Cooling
With Water Cooling (OS136-WC)
With Air Cooling (OS136-WC) 0 to 110°C (32 to 230°F)
Operating Relative Humidity: Less than 95% RH, non-condensing
Water Flow Rate for OS136-WC: 0.25 GPM, room temperature, minimum
Air Flow Rate for OS136-WC
Warm up Period:
Thermal Shock:
Air Flow Rate for Air Purge Collar 1 CFM (0.5 liters/sec.)
Transmitter Housing:
Power:
Dimensions:
Weight:
9
0 to 100ºC (32 to 212ºF)
-18 to 260ºC (0 to 500ºF)
-18 to 538°C (0 to 1000°F)
3% of Rdg or 4.4°C (8°F) whichever is greater
3% of Rdg or 5.5°C (10°F) whichever is greater
1% of Rdg or 1.0ºC (2.0ºF) whichever is greater
10 to 1
5.5 to 14 microns
150 msec, 0 to 63% of final value
0.5 to 1.0, adjustable via Single Turn Pot
Voltage, 100 mA Drive
0 to 100% Adj., set via Pot
4 to 20 mA
0 to 5 Vdc
0 to 10 Vdc
K type thermocouple, compensated
10 mV/ºC
10 mV/ºF
(0 to 5 Vdc) 1 K-Ohms
(0 to 10 Vdc) 1 K-Ohms
(4 to 20 mA) (Power Supply – 4)/20 mA
(10 mV/Deg) 1 K-Ohms
(K T/C) 10 K-Ohms
0 to 70ºC (32 to 158ºF)
0 to 200ºC (32 to 392ºF)
5 CFM (2.4 liters/sec)
1 to 2 minutes
About 30 minutes for 25ºC (77°F) abrupt
ambient temperature change
Stainless Steel 316, NEMA-4 & IP65 rated
12 to 24 VDC @ 50 mA
25.4 OD x 127 L mm (1.0" OD x 5.0" L)
0.80 lb (363 g)