Fixed alarms
High priority alarms:
Circuit disconnected—when leak exceeds
105 L/min (1.75 L/sec) for at least 15 sec
Over pressure
Blocked tube
Internal battery empty—when its charge level is
less than 15%, the system may stop operating in
a little as two minutes
Pressure sensor failure (system failure 7)
Motor failure (system failure 6)
Critical software failure (system failure 38)
Medium priority alarms:
High temperature (internal components such as
motor)
System failure (self test, flow sensor, software,
calibration)
Internal battery low—when its charge level falls
below 30%
Low priority alarms:
External DC power use
Internal battery use
Keypad failure
Finger sensor failure
System failure (software, system component)
Disconnected FiO
Disconnected XPOD pulse oximetry
High temperature attention (internal components
such as motor)
Resetting alarms
When the Alarm mute key is pressed during therapy for at least three seconds, the following
alarm messages and the audible alarm will be temporarily cleared:
High Pressure
Low Pressure
High Leak
Non-Vented Mask
Low Min Vent

Tailoring treatment setup options

Setting Ramp
If the patient experiences difficulty falling asleep with full pressure, they may wish to make use
of Ramp where the pressure starts very gently and slowly increases to the set treatment
pressure over a selected time period (see "Setup menu: Options" on page 42).
The Treatment screen following shows Max Ramp set at 15 minutes, while the graph shows
the pressures experienced during ramp while in bilevel mode.
CAUTION
EPAP/PEEP and Pressure Support achieve their configured values at the end of the ramp
period. Some patients require full support from the commencement of therapy. Ramp
may not be appropriate in such cases.
sensor
2
High Breath Rate
Low Breath Rate
High FiO
Low FiO
Apnoea
Low SpO
User adjustable alarms
High priority alarms:
Non-Vented Mask
Medium priority alarms:
Low Min Vent
High Leak
High Pressure
Low Pressure
High Respiratory Rate
Low Respiratory Rate
Apnoea
Low priority alarms:
High FiO
level
2
Low FiO
level
2
Low SpO
level
2
2
2
2
Starting therapy
25