Alarm
High Respiratory Rate
1. On the Setup, Clinical Settings menu, set Mode to ST and
Backup Rate to 30 bpm.
2. On the Setup, Alarm Settings menu, set All alarms off to Yes.
3. Now set High Resp Rate Alarm Setting to 10 bpm, and change the
alarm to be On.
4. Block the air tubing.
5. Start ventilation.
Low Respiratory Rate
1. On the Setup, Clinical Settings menu, set Mode to ST and
Backup Rate to 10 bpm.
2. On the Setup, Alarm Settings menu, set All alarms off to Yes.
3. Now set Low Resp Rate Alarm Setting to 30 bpm, and change the
alarm to be On.
4. Block the air tubing.
5. Start ventilation.
Low FiO
1. Connect FiO
2. Calibrate the FiO
3. On the Setup, Clinical Settings menu, set Mode to T.
4. Set Respiratory Rate to 10 bpm.
5. On the Setup, Alarm Settings menu, set All alarms off to Yes.
6. Now set Low FiO2 Alarm Setting to 80%, and change the alarm to
be On.
7 . Start ventilation.
High FiO
1. Connect the FiO
2. Calibrate the FiO
3. Set the High FiO
4. Start ventilation.
5. Start O
The alarm activates within 15 seconds.
High Leak
This test must be carried out using the patient's actual circuit and
interface (including catheter mount and cannula if in use). If oxygen is
in use, include the oxygen flow entrained to the oxygen inlet at the
rear of the device.
1. On the Setup, Clinical Settings menu, set Mode to T and
Respiratory Rate to 30 bpm.
2. Set the High Leak Alarm to ON.
3. Start therapy.
4. Partially block the air tubing.
The alarm is triggered when leak exceeds 40 L/min (0.67 L/sec) for
at least 20 sec. If too much leak occurs, the Circuit Disconnected
alarm may be activated.
Circuit Disconnected
1. Disconnect air tubing at the device air outlet.
2. Start therapy.
If your configuration does not adequately trigger this alarm, it is
due to the resistance of your accessories—not the device.
Consider using the High Leak, Low Minute Ventilation, Low
Pressure, Low Respiratory Rate and Apnoea alarms.
40
2
cell.
2
cell.
2
2
cell.
2
cell.
2
Alarm setting to 30%.
2
supply to provide higher FiO
2
2
level than alarm threshold.
Testing interval
Each new patient.
Each new patient.
Always when oxygen is
used and the FiO
cell
2
has been recalibrated.
Always when oxygen is
used and the FiO
cell
2
has been recalibrated.
On each new patient.
If required for the patient
particularly if the High
Leak alarm has been
disabled, to avoid
rebreathing if the air
circuit becomes
disconnected at the
device during the therapy.