Table of Contents

Advertisement

248476r4.book Seite 68 Freitag, 28. Oktober 2011 3:10 15
Recommended separation distances between portable and mobile RF communications equipment and the device
The device is intended for use in an environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the device as
recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of
transmitter (W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be determined using the equation
applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

Symbols

Follow instructions for use;
protected against solid foreign objects of 2.5 mm diameter and greater and vertically falling
water drops;
Connection for oxygen supply; max 30 l/min (max 30 l/min)
CE labeling in accordance with EC directive 93/42/EEC, class II b;
for storage and transport;
Manufacturer;
Do not re-use;
damaged;
LATEX
Environmental information This device must be disposed of in accordance with the laws and
regulations of the country in which disposal occurs.
The crossed-out wheeled bin symbol
be disposed of together with general household waste, but instead requires separate disposal.
This requirement for separate disposal is based on the European Directive 2002/96/EC for
electrical and electronic equipment, and the European Directive 2006/66/EC for batteries. You
can hand in the product at a municipal collection point, for example. This reduces the impact on
natural resources and prevents contamination of the environment through the release of
hazardous substances.
Batteries containing more than 0.0005 percent of mercury by mass, more than 0.002 percent
of cadmium by mass or more than 0.004 percent of lead by mass are marked below the
crossed-bin symbol with the chemical symbols (Hg, Cd, Pb) of the metals for which the limit is
exceeded.
For further information regarding product disposal, please contact your local ResMed office or
your specialist distributor, or visit our website at www.resmed.com.
Dispose of used air filters and air tubing according to the directives in your country.
68
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d = 1.17 √P
0.12
0.37
1.17
3.70
11.70
Class II equipment;
Caution;
Standby or preparatory state for a part of equipment;
Handle with care;
Upside;
Catalogue number;
Use by date;
Keep away from sunlight;
Latex-free
indicates that the product bearing this symbol may not
80 MHz to 800 MHz
800 MHz to 2.5 GHz
d = 1.17 √P
d = 2.33 √P
0.12
0.23
0.37
0.74
1.17
2.33
3.70
7 .37
11.70
23.30
Type BF equipment; IP31 Device is
Data communication port;
+ 60 °C
+ 140 °F
Temperature limitation
- 20 °C
- 4 °F
Maximum humidity;
Serial number;
Do not use if package is
Keep dry;
Batch code;

Advertisement

Table of Contents
loading

This manual is also suitable for:

C28554C29268C29718