Advanced Settings - Factory default settings and parameter ranges
Parameter
Min PS
Max PS
Rise Time
Fall Time
Ti Min
Ti Max
Trigger
36
Mode
Default
4 cm H
2
20 cm H
    
150
    
200
 
0.3 sec
 
2.0 sec
 
 
Medium
Description
O
Minimum pressure support in iVAPS mode.
Options: 0-20 cm H
O (Min PS always below
2
Max PS); 0.2 cm H
O increment
2
O
Maximum pressure support in iVAPS mode.
2
Options: 0-30 cm H
O; 0.2 cm H
2
Rise Time sets the time taken for the device to
pressurise to IPAP (see page 10).
Notes:
The Rise/Fall Time milliseconds scale is an
approximate only.
The actual Rise Time achieved is influenced by
factors including compliance, resistance, leak,
pressure differential and patient breathing
patterns.
Prolonged Rise Time inhibits pressurisation;
Care should be taken to ensure Rise Time is
set according to the patients inspiratory flow
and work of breathing. Patients with a high
work of breathing, short Ti and fast Respiratory
Rate will require shorter rise times than those
with lower Respiratory Rate, quiet breathing
and normal Ti.
Options: min, 150–900 (limited to 2/3 of Ti Max
or Ti); 50 increments
Fall Time sets the time taken for the device to
depressurise to EPAP/PEEP when cycling occurs
(see page 10).
Notes:
The Rise/Fall Time milliseconds scale is an
approximate only.
The actual Fall Time achieved is influenced by
factors including compliance, resistance, leak,
pressure differential and patient breathing
patterns.
Options: min, 100–400; 100 increments
Sets the minimum limit on the time the device
spends in IPAP (see page 9).
Options: 0.1 sec–Ti Max; 0.1 sec increments
Sets the maximum limit on the time the device
allows the patient to spend in inspiration (see
page 9).
Options: 0.3–4 sec or 2/3 of one breath cycle
defined by Backup Rate; 0.1 sec increments
Sets the sensitivity for triggering. Triggering
occurs when inspiratory flow increases above a
certain level and the device changes from EPAP
to IPAP (see page 9).
Options: very high, high, medium, low, very low
O increments
2