The alarm volume can be set Low, Medium or High. From the Setup menu, select Alarm
Settings. After the set value has been confirmed, the alarm will sound and the alarm LED lights.
Alarm settings, see "Setup menu: Alarm Settings (Alarm Volume)" on page 14.
You can mute an alarm by pressing
alarm sounds again. When an alarm is muted, the Alarm mute key LED will light constantly.
For a high or medium priority alarm, if after two minutes the problem is still present, the
alarm will sound again. Any active low priority alarm will be permanently muted and the
Internal Battery Use alarm will be cleared until the alarm condition is fulfilled again.

Tailoring treatment setup options

Setting Ramp
Programs
Note: Each program retains its own circuit configuration. When switching between
programs, ensure you use the correct circuit (breathing system) as 'learned' for that
program. If more than one program has been set up, follow the directions given by your
clinician for when and how each program should be used.
once. By pressing the Alarm mute key again, the
Ramp time
Program
Ramp time is a feature that can be enabled
by your clinician by setting a maximum ramp
time. Designed to make the beginning of
treatment more comfortable, ramp time is
the period during which the pressure
increases from a low start pressure to the
treatment pressure. See "Setup menu:
Options" on page 14.
Programs can be configured by your clinician
to provide you with multiple treatment
options. For example, a clinician can set up
programs for sleeping versus waking use or
for use during exercise or physiotherapy.
Programs save different patient circuit
configurations (Learn Circuit result) and
therapy and alarm settings.
The Stellar comes with one active program.
Your clinician can configure two programs. If
your clinician has selected dual, you can
choose the program to use on the Treatment
screen after stopping treatment. If only a
single program is selected, the option does
not display.