Installation, Mounting & Wiring
• Immediately discontinue use of the battery if the battery emits an unusual
• Do not place the battery in microwave ovens or high-pressure containers.
S2-M
S2-D&E

4.5 Mounting

The DTG-RTD100 can be mounted in one of two different ways depending
upon if you are using an external RTD sensor or if you ordered a model with the
integral probe assembly.
Models Without Integral Probe
A mounting bracket accessory is available, model number: DTG-RTD100-MB.
The mounting bracket attaches to the DTG-RTD100 meter and allows for wall
mounting.
4-3
smell, feels hot, changes color or shape, leaks or appears abnormal in any
other way.
Standard 1.5V C cell Alkaline batteries can be used as an alternative
to the standard Lithium batteries. You must first remove all batteries
from the gauge. Then configure jumper S2 with only position M
installed as indicated in Figure 4-3. Always remove all 3 of the S2
jumpers before changing configurations. Failure to properly configure
the jumpers can damage the gauge. Battery life is subject to usage
conditions. Battery life estimates are provided for Lithium batteries only
TB2
BATTERY
Figure 4-3. Battery Type Configuration
NOTE:
BATTERY
ALKALINE SETTING*
M
E
TB1
D
LITHIUM SETTING*
M
E
D