To Check Out The Oscilloscope - Agilent Technologies 54610B User's And Service Manual

Table of Contents

Advertisement

Service
Troubleshooting the Oscilloscope

To check out the oscilloscope

Is there an interface module connected to the oscilloscope?
1
If yes, do the following steps. If not, go to step 2.
Turn off the oscilloscope.
a
Remove the module.
b
Turn on the oscilloscope, then check for the failing symptom.
c
If the failing symptom disappears, replace the module. If not, go to
step 2.
Disconnect any external cables from the front panel.
2
Disconnect the power cord, then remove the cover.
3
Connect the power cord, then turn on the oscilloscope.
4
If the display comes on after a few seconds, (logo and copyright text,
followed by a graticule with text at top of the display) go to "To check the
LVPS," on page 3-35. If after checking the LVPS the voltages are within the
test limits, go to step 8. If not, go to step 6. If the display did not come on,
do the steps below.
Check the intensity knob to see setting to see if its set too low.
a
If there is still no display, disconnect the power cord.
b
Check all cable connections.
c
Go to "To check the LVPS," on page 3-35.
d
If the voltages are within the limits go to step 5. If not, go to step 6.