Performance Characteristics

XY Operation

XY Operation
Operating mode: X=Ch 1, Y=Ch 2, Z=Ext. Trigger
Z Blanking: TTL high blanks trace
Bandwidths: X-axis and Y-axis same as vertical system
Z-axis is dc to 100 MHz
Phase difference: ±3 degrees at 100 kHz

Display System

Display: 7-inch raster CRT
Resolution: 256 vertical by 500 horizontal points
Controls: Front-panel intensity control
Graticule: 8 × 10 grid or frame
Storage Scope: Autostore saves previous sweeps in half bright display and the
most recent sweep in full bright display. This allows easy differentiation of
current and historic information.