Performance Characteristics

Horizontal System

Horizontal System
Sweep speeds: 5 s/div to 1 ns/div main and delayed
Accuracy: ±0.01% of reading
Vernier (Both main and delayed sweep): Accuracy ±0.05% of reading
Horizontal resolution: 25 ps
1
Cursor accuracy
(∆t and 1/∆t):
±0.01% ±0.2% of full scale ±200 ps
Delay jitter: 10 ppm
Pretrigger delay (negative time): ≥10 divisions
Posttrigger delay (from trigger point to start of sweep):
The greater of 2560 divisions or 50 ms, but not to exceed 100 s.
Delayed sweep operation:
From 2 times up to 200 times main sweep
Delayed sweep can be as fast as 1 ns/div but must be at least 2 times
main sweep.
Horizontal modes: Main, Delayed (Alt), X-Y, and Roll
1
Tested, see "To verify horizontal ∆t and 1/∆t accuracy," on page 3-14.