Verifying Oscilloscope Performance
To verify horizontal ∆t and 1/∆t accuracy
In this test you verify the horizontal ∆t and 1/∆t accuracy by measuring the
output of a time mark generator with the oscilloscope.
Test limits: ±0.01% ±0.2% of full scale ±200 ps (same channel)
Equipment Required
Critical specifications
Stability 5 ppm after 1/2 hour
BNC, 3 feet
Connect the time mark generator to channel 1. Then, set the time
mark generator for 0.1 ms markers.
Setup the oscilloscope.
, then press the Default Setup softkey.
Press
a
Setup
Press
b
Autoscale
Set the time base to 20 µs/div.
c
Adjust the trigger level to obtain a stable display.
d
3-14
.
Recommended Model/Part
TG 501A and TM 503B