Communication errors may occur during calibration when using a slow communication Baud
rate. If you receive a Failed to get calibration data... message, try increasing the Baud rate
(115200 Baud is recommended). The Workhorse II uses a Hard break.
3. Click the Calibrate button.
4. Select Use Pitch/Roll?
Select No. This calibration requires two rotations (one for calibration and one for
verification).
Select Yes. This calibration requires eight rotations (four for calibration and four for
verification).
5. Click the Start Calibration button.
6. As you rotate the system, the bars will change color. The Blue bar indicates where you are in the
rotations.
Green – Good
Light Green – Acceptable
Yellow – Within parameters (one or two yellow bars for the entire rotation is OK)
Orange – Unacceptable - Rotate slower!
Red – Not measured
7. When the first rotation(s) are complete, click OK on the message box to continue with the verifica-
tion samples.
8. When the second rotation(s) are complete, click OK on the message box. The calibration error
should be less than 2 degrees.
Page 14
EAR99 Technology Subject to Restrictions Contained on the Cover Page.
WorkHorse Monitor/Sentinel Deployment Guide