Teledyne WORKHORSE Deployment Manual page 10

Monitor / sentinel
Hide thumbs Also See for WORKHORSE:

Advertisement

P/N 957-6285-00 (July 2023)
If you are unsure of the ADCP's baud rate, use Tools, Find ADCP. TRDI Toolz will try different
baud rates until it connects to the ADCP.
>{¡┐φ²∩╜╜²²jñ~ªñδgJ
Checking 115200 baud rate
==
WorkHorse II Broadband ADCP Version 7x.xx
Teledyne RD Instruments (c) 1996-2022
All Rights Reserved.
>
If you don't know what com port(s) were added when using a USB to serial adapter, use
Windows Device Manager
note the list of ports, reinsert the adapter and note the new port.
Quick Review
Verify the wakeup banner displays
Install the USB to Serial adapter Virtual
COM Port (VCP) driver
Page 8
EAR99 Technology Subject to Restrictions Contained on the Cover Page.
Checking 9600 baud rate
®
to determine the Com port. Remove the adapter, wait a moment,
If you are unsure of the ADCP's baud rate,
use Tools, Find ADCP. TRDI Toolz will try
different baud rates until it connects to
the ADCP.
The default factory set communications
settings for Sentinel is RS-232, 9600-baud, no
parity, 8 data bits and 1 stop bit. Monitor
systems use RS-422, 9600-baud, no parity, 8
data bits and 1 stop bit.
For help on using TRDI Toolz, click the
icon.
If you have any problems connecting, see
the Workhorse Operation manual,
Chapter 6.
The free FTDI driver download page is
available here:
WorkHorse Monitor/Sentinel Deployment Guide
https://ftdichip.com/drivers/