3. Enter the RE ErAsE command and then press the Enter key or click on Send. This command is case
sensitive.
>RE ErAsE erasing...
Recorder erased.
>
Use TRDI Toolz to download data from the CompactFlash card (see Downloading Data Files).
Once erased, data is not
card using Windows®. This may leave hidden files on the card. Always use the ADCP's RE
ErAsE command to format and delete data from the CompactFlash card recorder.
A recorder card with a file count more than 50 files will cause a noticeable delay between the
first ensemble and the Start Pinging (CS) command. If this delay is a concern, ensure that
unwanted files are cleared before starting the deployment.
The maximum data file size is 4GB.
Testing the ADCP
To run the Built-in tests:
1. Setup the communication parameters between TRDI Toolz and the ADCP.
2. Wake up the ADCP by pressing the
3. Enter the PA direct command to send to the ADCP and then press the Enter key or click on
Send.
If any of the tests fail, read Chapter 6 in the Monitor/Sentinel Operation Manual.
Testing the Sensors
To test the sensors:
1. Setup the communication parameters between TRDI Toolz and the ADCP.
2. Wake up the ADCP by pressing the
3. Enter the PC2 direct command to send to the ADCP and then press the Enter key or click on
Send.
Press any key to quit sensor display ...
Heading
Pitch
Roll
301.01°
-7.42°
-0.73°
300.87°
-7.60°
-0.95°
4. Use the PC2 test to display the sensor values. Rotate and tilt the system and verify the Pitch and
Roll sensor data is reasonable. Rotate the system clockwise and verify the heading increases. Vali-
date the accuracy at 0, 90, and 180 degrees. If the heading is off by more than 2 degrees, calibrate
the compass. If the Depth sensor is not zero, zero the pressure sensor.
5. If a sensor fails, contact TRDI Field Service.
See the WorkHorse Operation Manual, Chapter 5 for details on testing the sensors.
Page 12
EAR99 Technology Subject to Restrictions Contained on the Cover Page.
recoverable. Do not format or delete files from the CompactFlash
button.
button.
Up/Down
Attitude Temp
Up
24.35°C
Up
24.36°C
WorkHorse Monitor/Sentinel Deployment Guide
Ambient Temp
Pressure
22.97°C
0.0 kPa
22.97°C
0.0 kPa