Download Print this page

Deploying The Adcp - Teledyne WORKHORSE Deployment Manual

Monitor / sentinel
Hide thumbs Also See for WORKHORSE:

Advertisement

WorkHorse Monitor/Sentinel Deployment Guide

Deploying the ADCP

D
E P L O Y I N G T H E
Setting the ADCP clock
Testing the ADCP
Calibrating the Compass
Sending the commands to the ADCP
Pre-Deployment Checks
TRDI Toolz has a user-friendly interface for running the pre-deployment tests and setting the Moni-
tor/Sentinel real-time clock.
Refer to the Workhorse and Workhorse II Command Guide for a listing of all direct
commands and their format.
Setting the ADCP Clock
The real-time clock (date and time) within the Monitor/Sentinel maintains the correct time while system
power is removed. The clock is powered by a lithium battery.
To set the ADCP's clock to match the PC time:
1. Setup the communication parameters between TRDI Toolz and the ADCP.
2. Wake up the ADCP by pressing the
3. Click Tools, PC Time to ADCP.
4. TRDI Toolz will send the TS command to set the clock.
>TS 22/06/17 09:50:34
>
Format the CompactFlash Card Recorder
This section applies to the WHII electronics only. See How do I know if I have WH or WHII
electronics? to determine what electronics are installed in the system.
Use the RE ErAsE command to erase and format the CompactFlash memory card to the FAT32 format.
To format the CompactFlash memory
1. Setup the communication parameters between TRDI Toolz and the ADCP.
2. Wake up the ADCP by pressing the
A D C P
I N C L U D E S T H E F O L L O W I N G S T E P S
button.
card:
button.
EAR99 Technology Subject to Restrictions Contained on the Cover Page.
P/N 957-6285-00 (July 2023)
: