MicroDock II
Users Manual
Conflicts
If a conflict occurs, it must be resolved to continue the bump
check or calibration. The option is provided to manually
select a gas inlet or to accept the default inlet that the station
selects.
If one or more Conflicts display, complete the following
steps or press C cancel to quit the current bump/calibration.
Press C more to access the gas type/inlet screen
1.
to view which gas is configured for more than one
inlet.
The
icon displays on the far right side on the row
*
of the gas that is setup for more than one inlet. The
50
*
also indicates the inlet that the station has
selected as the optimum inlet.
Press C OK to accept the inlet that is selected and
2.
automatically resume the bump check or
calibration.
Note
If sel is not selected within 30 seconds, the station
automatically defaults to the inlet it has selected.
Or
Press C
or C
3.
to move the
that is in conflict.
Press C sel to activate the inlet field. The
4.
automatically displays on the left side of the inlet.
Press C
or C
5.
to scroll to the desired inlet
(1-5).
icon to the gas
*