The cursor automatically displays below the first
value of the month. Each value is selected,
changed, and bypassed individually.
Press C
2.
to toggle between 0 or 1. When the
correct value displays, press C sel to confirm the
change. The cursor then automatically moves to
the second value of the month.
Or
Press C sel to bypass the first value. The station
automatically retains the current value and
proceeds to the second value of the month.
3.
Repeat step #2 for the remaining time and date
values.
Press C exit to return to the user options menu.
4.
Press C exit again to return to the normal
5.
operating screen.
32

Inlet Setup

The Inlet Setup option is used to setup the specific gas
cylinders for bump checks and calibrations. This option is
used to
select the gas type(s)
enter the corresponding gas concentrations level(s), and
enter the gas cylinder lot # of each selected gas.
a Warning
Failed tests can result if the inlets are not setup
correctly.
For initial station activation, ensure the inlets are installed
correctly as follows:
Inlet 1 is the default connection for ambient air.
Inlet 2 is the default connection for four-gas mix, unless
otherwise specified at purchase.
Inlets 3-5 are designed connections for additional
gases. However, unless specified at the time of
purchase, inlets 3-5 are configured for ambient air.