A3 Output Level Adjustment

A3 Output Level Adjustment
The purpose of this procedure is to adjust the A3 output level and flatness.

Required Equipment

Power Meter
Power Sensor
Calibration Kit
7mm-N(f) Adapter

Procedure

Step 1. Run the adjustment program.
Step 2. Choose the A3 Adjustment in the Spot Adjustment.
It is possible to perform this item with the All Adjustment too.
Step 3. Connect the instrument as shown in Figure 3-12.
Figure 3-12
A3 Output Level Adjustment Setup (1/2)
136
Description
*1.The E9304A cannot be used with the 438A.
Recommended Model
438A, E4418A/B or E4419A/B
8482A+8481D or E9304A
16195B
11524A
*1
Chapter 3