E4991A Control Mode

It is necessary to set the E4991A in addressable only mode to perform the adjustments.
When your E4991A is set to system controller mode, set the E4991A in addressable only
mode by the following E4991A mouse procedure.
Step 1. Click
on the
System...
Step 2. Click
GPIB Setup Menu
Step 3. Select
Addressable Only
Step 4. Turn the E4991A's main power off and then back on again.
Or, set to addressable only mode by the following E4991A front panel procedure.
Step 1. Press
key on the front panel.
[System]
Step 2. Press
key on the front panel.
[Click]
Step 3. Press
key again, if system controller mode is selected in the
[Click]
Step 4. Turn the E4991A's main power off and then back on again.
In addition, you can confirm the E4991A GPIB address by the above procedure.
Chapter 3
Preparation for using the Adjustment Program
menu.
System
box.
in the
box.
Control Mode
Adjustment
box.
Control Mode
121