Users Manual
A low resistance reading is required to confirm a good
ground connection through the power cord. Refer to the
appropriate electrical safety standard for the specific limit
value to be followed.
Figure 8 shows the electrical connections between the
Analyzer and the DUT. Table 4 lists the abbreviations used
in the schematics and their descriptions.
14
Table 4. Schematic Abbreviations
Abbreviation
MD
Measuring Device (ESA609 Analyzer)
FE
Functional Earth
PE
Protective Earth
Mains
Mains Voltage Supply
L1
Live Conductor
L2
Neutral Conductor
DUT
Device Under Test
DUT_L1
Device Under Test Live conductor
DUT_L2
Device Under Test neutral conductor
DUT_PE
Device Under Test protective earth
REV POL
Reversed mains supply polarity
PE Open
Open protective earth
Test Voltage
Meaning