Turn on the Analyzer

To turn on the Analyzer, connect the power cord and plug
the cord into an outlet. The Analyzer will perform a series
of self-tests and then display the message shown in
Figure 5 when the self-test has completed successfully.
Figure 5. Start screen
During the self-test, the Analyzer checks its ac mains input
for proper polarity, ground integrity and voltage level. If the
polarity is reversed, the Analyzer indicates this condition. If
the ground is open, the Analyzer displays this fault. If the
mains voltage is too high or too low, the Analyzer displays
this fault and does not continue until the supply voltage is
corrected and the Analyzer power cycled off and then on
again.

Accessing the Analyzer's Functions

Use the test function buttons on the Analyzer to select the
type of test. Use the menus to select specific tests or setup
options. For example, press  and the Analyzer shows
the available leakage current tests at the bottom of the
display. Press a softkey (F1 through F3) under a specific
test to setup for or perform the selected test.
When an option, such as the Earth condition, is not
displayed, it cannot be changed. However, the Analyzer
indicates the setting.

Setting Up the Analyzer

Use the rocker switches to switch between normal and
reversed polarity.
fis201.bmp
Electrical Safety Analyzer
Turn on the Analyzer
11