Performing a Direct-Equipment Leakage Test

The Direct-Equipment Leakage test is available
when the EN62353 standard is selected.
The Direct-Equipment Leakage Current test measures the
leakage current between the exposed conductive surface
on the housing, to mains earth.
To perform a direct-equipment test, press .
The direct-equipment test is the default test and should
already be selected.
Figure 12 shows the electrical connections between the
Analyzer and the DUT during a Direct-Equipment Leakage
Current Test.
Note
The following outlet conditions apply when performing this
test:
Normal Polarity, Closed Earth
Normal Polarity, Open Earth
Reversed Polarity, Closed Earth
Reversed Polarity, Open Earth
Electrical Safety Analyzer
Performing Electrical Safety Tests
21