Replacing The Batteries; Errors And Troubleshooting; Care And Cleaning - Braun ThermoScan 7+ Owner's Manual

Ear thermometer
Hide thumbs Also See for ThermoScan 7+:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
How to set a default age group
You have the option to set a default Age Precision age group. This will allow the device to power
on and be used in default selection without re-selecting with each power on.
1. When the device is off, hold both the power button and the Age Precision button
for 4 seconds, the backlight will turn yellow, you can then release the buttons.
2. All Age Icons will flash continuously until the Age Precision button is pressed.
3. Press the Age Precision button, the 0-3 month icon will be filled in and flash and the other
icons will be empty. A second button press will advance to the 3-36 month icon flashing, a
third will advance to 36+months flashing. This cycle continues in order with each button press.
A full beep will sound with every button press.
4. After 3 seconds of no button press, the displayed icon will remain on the screen and a full beep
will sound. The device will turn off. When the device is turned back on, the default will be set.
Default will be set on the device until the batteries are removed and then put back in.
How to change temperature between °C and °F
Your Braun ThermoScan is shipped with the Celsius (°C) temperature scale activated. If you wish to switch to Fahrenheit
(°F) and/or back from Fahrenheit to Celsius, proceed as follows:
1. Make sure the thermometer is turned off.
2. Press and hold down the Power button
. After about 3 seconds the display will show this
sequence: °C / SET / °F / SET.
3. Release the Power
button when the desired temperature scale is shown. There will be a
short beep to confirm the new setting, then the thermometer is turned off automatically.

Care and cleaning

The probe tip is the most delicate part of the thermometer. It must be clean and intact to ensure accurate
readings. If the thermometer is ever accidentally used without a Hygiene cap, clean the probe tip as follows:
Very gently wipe the surface with a cotton swab or soft cloth moistened with alcohol. After the
alcohol has completely dried out, you can put a new Hygiene cap on and take a temperature
measurement.
If the probe tip is damaged, please contact Consumer Relations.
Use a soft, dry cloth to clean the thermometer display and exterior. Do not use abrasive cleaners.
Never submerge this thermometer in water or any other liquid. Store thermometer and Hygiene
caps in a dry location free from dust and contamination and away from direct sunlight.
Additional Hygiene caps (LF 40) are available at most stores carrying Braun ThermoScan.
6

Replacing the batteries

1.
Insert new batteries when the battery symbol appears on the display.
4 sec
2. Open the battery compartment. Remove the batteries and replace with new batteries, making
sure the poles are in the right direction.
3. Snap battery door into place.
To protect the environment, dispose of empty batteries at appropriate collection site
according to national or local regulations.

Errors and troubleshooting

Error Message
3 sec
Situation
Solution
No Hygiene cap is attached.
Attach new, clean Hygiene cap.
The thermometer probe was not positioned
Take care that the positioning of the probe is correct
securely in the ear.
and remains stable.
An accurate measurement
was not possible.
Replace Hygiene cap and reposition.
Press Start button to begin a new measurement.
ERROR + position error animated icon
Ambient temperature is not within
Allow the thermometer to remain for
the allowed operating range
30 minutes in a room where
(10 – 40 °C or 50 – 104 °F).
the temperature is between 10 and 40 °C
or 50 and 104 °F .
Temperature taken is not within
Make sure the probe tip and lens are clean and a
typical human temperature range
new, clean Hygiene cap is attached. Make sure the
(34 – 42.2 °C or 93.2 – 108 °F).
thermometer is properly inserted. Then, take a new
temperature.
HI = too high
LO = too low
System error – self-check display flashes
Wait 1 minute until the thermometer turns off
continuously and will not be followed by
automatically, then turn on again.
the ready beep and the ready symbol.
If error persists,
... reset the thermometer by removing the batteries
and putting them back in.
If error still persists,
... please contact the Service Centre.
Battery is low, but thermometer will still
Insert new batteries.
operate correctly.
Battery is too low to take correct
Insert new batteries.
temperature measurement.
System error without messages or
Insert new batteries. If the error persists, call
completely dead batteries
customer service.