How to set a default age group
You have the option to set a default AgeSmart™ age group . This will allow the device to power on
and be used in default selection without re-selecting with each power on .
1 . When the device is off, hold both the power button and the AgeSmart™ button for 4 seconds,
the backlight will turn yellow, you can then release the buttons .
2 . All Age Icons will flash continuously until the AgeSmart™ button is pressed .
3 . Press the AgeSmart™ button, the 0-3 month icon will be filled in and flash and the other icons
will be empty . A second button press will advance to the 3-36 month icon flashing, a third will
advance to 36+months flashing . This cycle continues in order with each button press . A full
beep will sound with every button press .
4 . After 3 seconds of no button press, the displayed icon will remain on the screen and a full beep
will sound . The device will turn off . When the device is turned back on, the default will be set .
4 sec
Default will be set on the device until the batteries are removed and then put back in .
How to change temperature between °C and °F
Your Braun ThermoScan® is shipped with the Celsius (°C) temperature scale activated . If you wish to switch to
Fahrenheit (°F) and/or back from Fahrenheit to Celsius, proceed as follows:
1 . Make sure the thermometer is turned off .
2 . Press and hold down the Power button
. After about 3 seconds the display will show this
sequence: °C / SET / °F / SET .
3 . Release the Power
button when the desired temperature scale is shown . There will be a
short beep to confirm the new setting, then the thermometer is turned off automatically .

Care and cleaning

The probe tip is the most delicate part of the thermometer . It must be clean and intact to ensure accurate
readings . If the thermometer is ever accidentally used without a lens filter, clean the probe tip as follows:
Very gently wipe the surface with a cotton swab or soft cloth moistened with alcohol . After the
alcohol has completely dried out, you can put a new lens filter on and take a temperature
measurement .
If the probe tip is damaged, please contact local service center .
Use a soft, dry cloth to clean the thermometer display and exterior . Do not use abrasive cleaners .
Never submerge this thermometer in water or any other liquid . Store thermometer and lens filters
in a dry location free from dust and contamination and away from direct sunlight .
Additional lens filters (LF 40) are available at most stores carrying Braun ThermoScan® .
6
4 sec
4 sec
4 sec
3 sec

Replacing the batteries

1 .
Insert new batteries when the battery symbol appears on the display .
3 sec
2 . Open the battery compartment . Remove the batteries and replace with new batteries, making
sure the poles are in the right direction .
3 . Snap battery door into place .
To protect the environment, dispose of empty batteries at appropriate collection site
according to national or local regulations .
3 sec

Errors and troubleshooting

Error message
Situation
3 sec
No lens filter is attached .
The thermometer probe was not positioned
securely in the ear .
An accurate measurement
was not possible .
ERROR + position error animated icon
Ambient temperature is not within
the allowed operating range
(10 – 40 °C or 50 – 104 °F) .
Temperature taken is not within
typical human temperature range
(34 – 42 .2 °C or 93 .2 – 108 °F) .
HI = too high
LO = too low
System error – self-check display flashes
continuously and will not be followed by
the ready beep and the ready symbol .
If error persists,
If error still persists,
Battery is low, but thermometer will still
operate correctly .
Battery is too low to take correct
temperature measurement .
System error without messages or
completely dead batteries
English
Solution
Attach new, clean lens filter .
Take care that the positioning of the probe is correct
and remains stable .
Replace lens filter and reposition .
Press Start button to begin a new measurement .
Allow the thermometer to remain for
30 minutes in a room where
the temperature is between 10 and 40 °C
or 50 and 104 °F .
Make sure the probe tip and lens are clean and a
new, clean lens filter is attached . Make sure the
thermometer is properly inserted . Then, take a new
temperature .
Wait 1 minute until the thermometer turns off
automatically, then turn on again .
. . . reset the thermometer by removing the batteries
and putting them back in .
. . . please contact the Service Centre .
Insert new batteries .
Insert new batteries .
Insert new batteries . If the error persists, call
customer service .