Table 10-7
Cable gland
M20 x 1.5
M20 x 1.5
M20 x 1.5
M25 x 1.5
Restraining capacity in accordance with DIN EN 50262 Class B
IP55 protection is only achieved if the diameters of the inserted cables are within the clamping range of
the cable gland.
10.3

OXYMAT 7 technical specifications

OXYMAT 7
The information on the smallest measuring span is available on the module nameplate.
Table 10-8
General
Information on the smallest span
Power consumption
Weight of analyzer section
Electrical safety
Table 10-9
Measuring ranges
Number of measuring ranges
Parameters can be as‐
signed in the measuring
ranges
Table 10-10
Gas inlet conditions
Sample gas pressure
Field device
Operating Instructions, 05/2022, A5E35640463-07
Field control unit: Cable gland
General technical specifications
Warm-up phase
Operating phase
Measuring ranges
Smallest possible spans
Largest possible span
Gas inlet conditions
for non-combustible
gases
for combustible gases
up to gas mixtures
which are occasionally
explosive
10.3 OXYMAT 7 technical specifications
Clamping range diameter
6.1 to 11.5 mm
9.1 to 15.9 mm
12.5 to 20.9 mm
14 to 22 mm
See module rating plate or device certificate
≤ 95 W
≤ 35 W
Approx. 5.5 kg (standard version)
In accordance with DIN EN 61010-1
Max. 4; parameters can be assigned freely
0.5 %, 1%, 2 % or 5 % O
100% O
2
500 to 2500 hPa (abs.)
800 ... 1100 hPa (abs.)
Technical specifications