Oxymat 7 Technical Specifications - Siemens SIPROCESS GA700 Compact Operating Instructions

Continuous gas analysis, devices in explosion-proof design
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Available languages

Available languages

Table 8-10
Option modules
Option module 1.1
Option module 2.1
Option module 2.2
Table 8-11
Climatic conditions
Permissible operating height
Permissible ambient temp. during storage and
transport
Ambient humidity (rel. humidity) during storage,
transport or operation
Permissible ambient temperature and humidity
8.2.4

OXYMAT 7 technical specifications

Table 8-12
General
Information on the smallest span
Power consumption
Weight
Electrical safety
Devices in explosion-proof design
Compact Operating Instructions, 05/2015, A5E35134047-02
Wall housing: Option modules
Wall housing: Climatic conditions
OXYMAT 7: General technical specifications
Warm-up phase
Operating phase
Technical specifications
8.2 General technical specifications
12 digital outputs, with changeover contacts, max.
load: 24 V AC/DC/1.7 A (total load for all 12 relay
outputs in continuous operation max. 244 W), float‐
ing, non-sparking
8 digital inputs, designed for 24 V, isolated, freely
parameterizable
6 analog outputs, 0/4 to 20 mA, floating; maximum
load 750 Ω
6 digital outputs, max. load: 24 V AC/DC/1.7 A (to‐
tal load for all 6 relay outputs in continuous opera‐
tion max. 122 W), floating, non-sparking
6 analog outputs 0/4 to 20 mA, isolated; max. load
750 Ω
4 analog inputs 0/4 to 20 mA, non-isolated, internal
resistance 100 Ω
4 digital inputs, designed for 24 V, isolated
3 000 m above sea level
-30 ... +65 °C
< 90%
→ Technical specifications for operation in hazard‐
ous areas (Page 88)
See module nameplate or device certificate
≤ 60 W
≤ 35 W
< 5.5 kg (standard version)
In accordance with DIN EN 61010-1
83