Siemens SIPROCESS GA700 Operating Instructions Manual page 74

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Technical specifications
10.2 Technical specifications of field device
Table 10-3
Electrical characteristics
Auxiliary power supply
Power consumption
EMC interference immunity (electro‐
magnetic compatibility)
Electrical safety
Power supply unit cable
Table 10-4
Electrical inputs and outputs
Digital outputs (DO)
Digital inputs (DI)
Standard Ethernet interface (bottom)
Service interface (inside at PU)
Table 10-5
Option modules
Optional module 2.2
Table 10-6
Climatic conditions
Permissible operat‐
ing elevation
Permissible ambient temperature
Permissible humidity
Under atmospheric conditions at 0.8 to 1.1 bar
1)
74
Field control unit: Electrical characteristics
Field control unit: Electrical inputs and outputs
Field control unit: Option modules
Field control unit / field module Climatic conditions
Hazardous areas:
1)
Non-hazardous
areas:
100 to 240 V AC (nominal range of use 85 to 264 V)
50 to 60 Hz (nominal range of use 47 to 63 Hz)
280 VA max.
In accordance with the standard requirements of NA‐
MUR NE21 (05/2006) and EN 61326-1 (01/2008)
In accordance with EN 61010-1, overvoltage category II
Max. diameter 10 mm, max. 3 x 2.1 mm
5, with changeover contacts, can be freely parameterized,
e.g. for measuring range identification; max. load: 24 V AC/DC/
1.7 A, isolated, non-sparking
5, designed for 24 V, isolated, freely parameterizable, e.g. for
autoranging
Ethernet RJ 45, 100 MBit
Ethernet RJ 45, 100 Mbit
3 analog outputs 0/4 to 20 mA, isolated; max. load 750 Ω
4 analog inputs 0/4 to 20 mA, non-floating, internal resist‐
ance ≤ 100 Ω
2000 m AMSL to EN 60079-0
3000 m above sea level
-30 to +70 °C during storage and transportation
+5 °C to +55 °C for regular operation with OXYMAT 7
+5 °C to +60 °C for operation with OXYMAT 7 and with
limited accuracy
< 90% relative humidity as annual average, during storage and
transportation (must not fall below dew point)
Operating Instructions, 05/2022, A5E35640463-07