Field Control Unit; Notes On Selection Connection Cables; Terminal Block - Siemens SIPROCESS GA700 Operating Instructions Manual

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Wiring

6.5 Field control unit

See also
Field control unit (Page 48)
6.5
Field control unit
6.5.1

Notes on selection connection cables

Note
Terminal ranges of the cable glands
Use cables matched to the terminal ranges of the cable glands:
• Cable gland M25 x 1.5 (nickel-plated brass), terminal range 14 ... 22 mm
• Cable gland M20 x 1.5 (nickel-plated brass), terminal range 6.1 ... 11.5 mm
• Cable gland M20 x 1.5 (nickel-plated brass), terminal range 9.5 ... 15.9 mm
For additional information, see → Technical specifications (Page 73).
IP55 can only be guaranteed if you use cables with the appropriate properties.
Note
Strain relief
The cable gland for strain relief has a restraining capacity in accordance with DIN EN 50262 Class
B.
If necessary, provide an external strain relief.
6.5.2

Terminal block

Overview
All electric connections and also the signal connections of the field control unit are present on
the terminal block on the DIN rail in the field control unit. Also connect the cable coming from the
field module here. The cable connections on the device side of the field module are already
preassembled when delivered.
Note
Strain relief of the connecting cables
Make sure cables are routed without force so that the strain relief of the cable gland is not
overloaded.
If necessary, install an external strain relief.
48
Operating Instructions, 05/2022, A5E35640463-07
Field device

Advertisement

Table of Contents
loading

Table of Contents