Oxymat 7 Technical Specifications - Siemens SIPROCESS GA700 Compact Operating Instructions

Continuous gas analysis, devices in explosion-proof design
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Table 10- 11 Wall housing: Option modules
Option modules
Option module 1.1
Option module 2.1
Option module 2.2
Table 10- 12 Wall housing: Climatic conditions
Climatic conditions
Permissible operating altitude
Permissible ambient temperature dur-
ing storage and transport
Ambient humidity (rel. humidity) during
storage, transport or operation
Permissible ambient temperature and
humidity
10.2.4

OXYMAT 7 technical specifications

Table 10- 13 OXYMAT 7: General technical specifications
General
Information on the smallest
span
Max. power
consump-
tion
Weight (only module)
Electrical safety
Max. EMC error during op-
eration with one and/or two
OXYMAT 7/ ULTRAMAT 7
Devices in explosion-proof design
Compact Operating Instructions, 05/2018, A5E35134047-03
OXYMAT 7
Warm-up
≤ 60 W
phase
< 5.5 kg
In accordance with EN 61010-1
10.2 General technical specifications
12 digital outputs, with changeover contacts, max. load: 24
V AC/DC/1.7 A (total load for all 12 relay outputs in continu-
ous operation max. 244 W), floating, non-sparking
8 digital inputs, designed for 24 V, isolated, freely parame-
terizable
6 analog outputs, 0/4 to 20 mA, floating; maximum load
750 Ω
6 digital outputs, max. load: 24 V AC/DC/1.7 A (total load
for all 6 relay outputs in continuous operation max. 122 W),
floating, non-sparking
6 analog outputs 0/4 to 20 mA, isolated; max. load 750 Ω
4 analog inputs 0/4 to 20 mA, non-isolated, internal re-
sistance 100 Ω
4 digital inputs, designed for 24 V, isolated
3 000 m above sea level
-30 ... +65 °C
< 90%
Technical specifications for operation in hazardous areas
(Page 97)
See module nameplate or device certificate
1.0% in accordance with EN 61326-1 (2013)
Technical specifications
OXYMAT 7 high temperature
≤ 120 W
87