When a dispensing pump is selected, the following will appear in the
display:
Dispenser calibration
V:
100 ml
Tip: The dispensing quantity can be set from 0 to 9999 ml in 1 ml in-
crements.
Press the multifunction button
The dispensing hose will fill up.
Stop the process by pressing the
Run the hose of the dispensing pump upwards into a measuring
vessel.
Hold the entire measuring vessel above the dispensing pump while
doing this.
Irritant detergents.
Skin and eye irritation.
Avoid contact with skin and eyes, wear protective clothing.
Return to the main menu by pressing the "Back" button.
Dispenser calibration
Dispenser pump 1
Dispenser pump 2
Select a dispensing pump.
Dispenser calibration
V:
100 ml
Press the multifunction button
Stop the pump by pressing the
automatically after 60 seconds.
Measure the required amount in the measuring vessel.
Dispenser calibration
V:
100 ml
Enter the amount measured in the
P
The throughput
is calculated in
Dispenser pump 1
s
P:
No
t:
0
0/1
0/1
.
0/1
Main menu
Dispenser pump 1
t:
0
s
P:
No
0/1
0/1
.
0/1
button, or the pump will stop
Dispenser pump 1
t:
s
P:
200,0 ml min
30
0/1
V:
field.
ml/min
Optional modules
button.
/
by the controls.
51