CALIBRATION CONSTANTS
SERIAL NUMBER; 0000
834041 BREV
01
AUG 86
1.
DWELL
TIME
SO
II
34.
AT90
SLOPE
-10
11
67.
AT60
SLP
20GZ
10
2.
YTMBXDLY2
74
II
35.
LEV
DAC
OFF; SYS
-2
II
68.
AT70
SLP
20GZ
13
3.
YTMBXDLY3
69
II
36.
LEV
DAC
GAIN;
SYS
-27
II
69.
AT80
SLP
20GZ
15
4.
YTMBXDLY4
100
11
37.
ADC
OFF SYS
-1
II
70.
AT90
SLP
20G2
18
5.
YTMDLY1
80
II
38.
ADCGAINSYS
-3
II
71.
YTM
OFFSET
1
1055
6.
YTMDLY2
101
II
39.
ADC
GAIN
LO
-40
II
72.
YTM
OFFSET
2
1108
7.
YTMDLY3
100
II
40.
ADCGAINHI
-40
II
73.
YTM
OFFSET
3
902
8.
YTMDLY4
100
II
41.
MAX SWEEP
RATE
600
II
74.
YTM
OFFSET
4
1024
9.
YTMGAIN1
1084
II
42.
ADC
OFFSET
3
II
75.
YTM SGL BD
RT
25
10.YTMGAIN2
964
II
43.AMOFFSET
2
II
76.
YTM MLT BD RT
1000
11.
YTM
CAIN
3
1151
II
44.
LVL
DAC
OFF; INT
-6
II
77.
YTM
BX
RTA
-25
12.
YTM
GAIN
4
1024
11
45.
LVL
DAC
OFF; EXT
0
11
78.
YTM BX RTB
25
13.
LO
SLOPE
49
II
46.
LVL
DAC
GAIN;
LO
11
11
79.
01 MODEL
5
14.
HI
SLOPE
117
II
47.
LVL
DAC
GAIN; HI
12
II
80.
YTM TEMP COMP
-17
15.
9G2
SLOPE
80
II
48.
LVL
DAC
GAIN;
EXT
-23
II
81.
LOCKMORD
0
16.
20GZ SLOPE
129
11
49.
PMRSMPGAIN
4
11
82.
LOCKMORD
ENABLE
O
17.
AT10
OFFSET
-5
11
50.
YTMBXGAIN1
0
II
83.
unused
0
18.
AT20 OFFSET
-7
II
51.
YTM BX
GAIN
2
0
11
84.
unused
0
19.
AT30 OFFSET
-5
11
52.
YTM
BX
GAIN
3
0
11
85.
unused
0
20.
AT40
OFFSET
-9
11
53.
YTM BX
GAIN
4
0
11
86.
unused
0
21.
AT50 OFFSET
-10
II
54.
STOP
LIMIT
20000
II
87.
unused
0
22.
AT60 OFFSET
-11
II
55.
START
LIMIT
10
II
88.
unused
0
23.
AT70 OFFSET
-14
II
56.
PRESET OPTION
O
II
89.
unused
0
24.
AT80 OFFSET
-20
II
57.
HPIBADDRESS
19
II
90.
unused
0
25.
AT90 OFFSET
-23
II
58.
RETRACE DWELL
O
II
91.
unused
0
26.
AT10
SLOPE
-1
II
59.
ATTENCONFIG
9716
II
92.
unused
0
27.
AT20
SLOPE
-2
II
60.
CONFIGURATION
513
II
93.
unused
0
28.
AT30
SLOPE
-3
II
61.
SERIAL"
378
II
94.
unused
0
29.
AT40 SLOPE
-5
II
62.
AT10 SLP 20G2
4
II
95.
unused
0
30.
AT50 SLOPE
-S
II
63.
AT20 SLP 20GZ
5
II
96.
unused
0
31.
AT60
SLOPE
-7
II
64.
AT30
SLP
20GZ
5
II
97.
unused
0
32.
AT70
SLOPE
-8
II
65.
AT40
SLP
20GZ
8
II
98.
MODEL'
5
33.
AT80
SLOPE
-9
II
66.
AT50 SLP 20GZ
9
II
99.
CHECKSUM
24097
Procedure for manually entering calibration
data into the HP
8340/41
B:
1.
Push the following sequence of KEYS:
Instrument Preset
SHIFT MHz
1 2
Hz
SHIFT
KHz
2 2
Hz
2.
Note the Entry Display
will
indicate the Calibration Constant number and value.
3.
Enter via the KEY BOARD or DATA KNOB the correct value for the
first
Calibration Constant indicated
in
the
display.
«
Terminate KEYBOARD
entries with the
Hz
key
»
4.
Go to the next Calibration Constant by pushing the UP step key. The next constant can then be entered. Do no
enter
the
"CHECKSUM"
Constant. (This
is
computed
automatically)
5.
The Step Keys can be used to move from one Calibration Constant to another to either check them or to correc
them.
6.
After
all
entries have been
made, check that
all
numbers are correct by using the step keys to review and
verify
them.
7.
Allow instrument
to warm
up for 1/2 hour and
make
sure that nothing
is
connected
to the Stop
Sweep connecto
on the
rear
panel. Push
SHIFT
PEAK to perform an Automatic Tracking Calibration. This step may modify
the
"YTM
GAIN n"
&
"YTM BX GAIN
n" consts.
8.
The CALIBRATION data should be permanently stored
in
the Non Volatile Protected Memory by pushing
the
following key
sequence:[SHIFT] [MHz]
[14]
[Hz] [SHIFT] [KHz] [5349] [Hz] [PRESET].
i
t
t
y
r
i
3
Figure A-3.
Sample Calibration
Constants Printed Copy (Found Inside Instrument)
A
-24
Assembly-Level Service Calibration constants