Sometimes Keystrokes are Missed
by
the Instrument
1.
The problem
is in
the A6 key down timer.
The instrument Responds
Only to
Keystrokes and/or the Front Panel
Knob
when
in
a
Swept Mode
1.
The
problem
is in
the A6
SRQ
buffer.
No
Response
when the Front
Panel Knob
is
Turned
1.
Press
a
function key and verify
that the ENTRY ON LED
lights.
If
this
LED does
not
light,
the
trouble
is
the annunciator latches.
2.
If the
ENTRY
ON LED
is
on, probe CLK (A6TP1) with an oscilloscope set
to 5
ms/Div. You
should
see
low going pulses as you turn the front panel knob, and the pulse width should vary with the
speed
of rotation.
If
CLK
is
bad, the trouble
is in
the
RPG (rotary
pulse generator) or the
RPG
counters/data
buffers.
3.
Probe
HI
RPG
SRQ
(A6TP8). You should
find high going pulses
that vary
in
width depending on
the response time of the instrument processor to
an
SRQ. These
pulses should be between 100
US
and
25
ms.
If
HI
RPG SRQ
is
correct, the trouble
is
the
SRQ
buffer or the RPG counters/data
buffers.
If HI
RPG
SRQ
is
not correct, the trouble
is
the
RPG count window timer.
Turning
the
RPG Causes Either
Very Small
or
Very Large Changes
in
the
Active Function
in
the ENTRY DISPLAY
1.
The
problem
is in
the A6 RPG count window timer.
The Instrument Does Not Preset when [INSTR PRESET]
is
Pressed
1.
The problem
Is in
the
A6 instrument preset
buffer.