15.
Refer to
Figure
B-24, Reference
Loop
-
M/N
Loop Assemblies and Mounting Hardware. On the
bottom of the instrument, remove
all
cables going
to the
A29, A30,
A31,
and
ASS
assemblies
16.
Again refer
to
Figure B-24. With
a
small pozidnv,
remove
all
28
screws that hold the A29, A30, A31,
and A33 sheet metal covers
in
place. Note that
3
screws are
shown
in
Figure B-24 which do
not
need to be removed. Remove the A29, A30,
A31, and
A33
assemblies.
Do Not
Remove
Figure B-24. Reference
Loop
-
M/N
Loop Assemblies and Mounting
Hardware
HP 8340B/41B
Reference
Loop
-
M/N
Loop
Repair Procedures
B-47