Transformer Installation
1.
Reverse the above procedure.
NOTES:
When
you reverse step 4a, do not allow the transformer secondary wires to touch
adjacent
A62 motherboard
lugs.
Ensure that the white transformer secondary wire pair
is
reattached
to the A62 motherboard lug
labeled
(9).
DO NOT attach
it
to
the
taller,
unmarked
lug.
When reversing step 4a, ensure the
blue
transformer secondary
wires are completely inside the
plastic clamp.
If
the plastic clamp
hardware
is
mounted
in
the wrong side panel
hole, the
A19 assembly
will not
go back
in
place.
Reinstall the A19 capacitor assembly as follows:
1.
Lift
the
two grey cables out of the way.
2.
Lower
the A19 assembly flange-end first
until
the finger contact end drops
in
place.
3.
Lift
up slightly on
the flange-end, and push
down on
the finger
connector side
until
the fingers
are
fully seated.
LINE MODULE
REPLACEMENT
Read the entire procedure before
you start!
Line Module Removal
Refer
to Figure
1-6.
1.
Ensure that the
main line fuse and the selected line
voltage are
correct.
2.
Disconnect
line
power,
and
remove the instrument
top and bottom covers.
3.
Wait
for the A19
power-on safety
indicator
to
go
out.
4.
Remove
the portion of the
rear
panel
that
holds the line module:
a.
Remove two screws
(item
12).
b.
Invert
the instrument and remove three rear
panel
screws (item
7).
c.
Remove the A35 assembly
protective cover
d.
Remove three rear panel screws
(item 13).
e.
Remove
the fan.
f.
Remove three rear panel screws (item
8).
g.
From
inside the instrument, push the rear
panel out.
5.
Desolder
all
wires from the
line
module.
HP8340B/41B
Change
4
Power
Supplies/Fan Repair
Procedures
1-19