Numeric Displays are Blank
1.
Connect the
oscilloscope to the CLK test point
(A2TP6), and set
it
to
0.5 ns/div,
1
V/div (0V
=
center screen). The waveform displayed
should be low
for 2.5 to 3.0 (is, with
levels
between
+2V
and approximately
2.5V.
If this
signal
is
not present, the trouble
is
the
NUM
CLK on the A2
assembly.
2.
Troubleshoot
the A2 assembly.
Entry Display Is Blank
1.
Connect oscilloscope
to
ALPHA CLK
(A2TP2), and set
it
to 0.5 ^S/div, 1V/div
(OV=Center
screen).
The waveform
displayed should be
TTL
low-going pulses, 0.3
nS
wide, with
a
period of
1.4 iis.
If
this signal
is
not present, the trouble
is
the alpha display clock control
in
the A3 assembly.
2.
Connect
the oscilloscope to
DATA (A2TP1), set
as
in
step
1.
The waveform
display should
indi¬
cated
both high and
low signal
levels
approximately
1.4 (is
wide.
If
the
DATA
signal
is
always
low,
all
LEDs
in
the
ENTRY
display will be off.
If this
signal
is
not present, the trouble
is
the A3 assembly
alpha display
shift register.
t
3.
Troubleshoot the A2 assembly.
The Same Segment(s)
is
Missing
In
Each Numeric Digit
1.
Troubleshoot the A2 assembly.
The Same Column
is
Missing
in
Each ENTRY DISPLAY Character
1.
Troubleshoot
the A2 assembly.
One Entire
Numeric Display
is
Blank
1.
Troubleshoot
the A2 assembly.
The Display Goes
into
Self Test When the Instrument
is
Turned ON and
Does Not Respond to Normal Instrument Functions
1.
The
trouble
is
the
A3
instrument bus interface.
HP 8340B/41B
A3
Display
Processor
Assembly-Level Troubleshooting
G-9