5-16.
LEVELED RF OUTPUT ADJUSTMENTS
Reference
Performance
Test:
Maximum
Leveled Output
Power
and
Accuracy
Service Section:
RF Section
Description
In
this
procedure, the RF output
signal
is
checked
in
leveled mode.
An
oscilloscope and detector are
used to adjust the Step
Recovery Diode (SRD,
internal to the
SYTM) for
optimum
bias and to eliminate
the possible occurrence of "squegging".
In
the next
procedure,
the
ALC
loop gain
in
each band
is
adjusted for optimum operation-
A
final section
checks the RF output
signal with
a
Spectrum Analyzer
for the
appearance of squegging
in
any
of the
bands
and allows
a
last fine adjustment to eliminate any squegging that
is
observed.
DIGITAL
VOLTMETER
OUT
Z-AXIS
BLANK/MKRS
^
A26TP3
(MOD
LVL)
A24TP12
(SRO)
STOP SWEEP
f
IN/OUT
OSCILLOSCOPE
10:1 DIVIDER
PROBE
CHAN
A
|
CHAN
B
Figure 5-52.
Leveled Power
SRD
Bias Adjustment Test Setup
HP 8340B/41B
Adjustments