N NOTE N
For parallelism adjustment purposes, apply a 300 mm scale to the guide plate trailing end
(toward the IP removal unit) and measure the gap between the scale and guide plate.
Guide plate
(4) Complete installation by reversing the removal procedure.
N NOTE N
When installing the SK2 assembly, perform its specified adjustment procedure.
009-058-00
10.20.2000 FM2732 (2)
0.5 mm or greater gap = NG
Gap smaller than 0.5 mm = OK
300 mm scale
Less than 0.5 mm
"4.9 Cassette IN Sensor (SK2)"
CR-IR347 Service Manual
300 mm scale
MC - 61
FR7H3342.EPS