10.14 Mirror Driving Motor (MZ4)

I Removal procedure
(1) Remove the light-collecting guide assembly (for front surface).
"10.9 Light-Collecting Guide Assembly (for Front Surface)"
(2) Remove the scanning optics unit.
"9.4 Removing and Installing the Scanning Optics Unit"
(3) Remove the MZ4.
Ring
009-058-00
10.20.2000 FM2732 (2)
FR7H3080.EPS
#3
[Loosen] TWP3x6
(Loktite)
Cam
#4
[Remove] BR3x8 (x3)
CR-IR347 Service Manual
Subscanning unit
MZ4
#1
[Disconnect] FASTON terminal (+), red
#2
[Disconnect] FASTON terminal (-), black