I Installation procedure
N INSTRUCTION N
When mounting the cam on the MZ4, secure it with the MZ4 motor shaft (flat-machined
surface) positioned as indicated below:
N NOTE N
When connecting the FASTON terminals (+ and -), insert them all the way into their
positions.
REFERENCE
The FASTON terminal (+/-) connectors are marked "+" or "-". Note the "+" and "-"
markings and make proper terminal connections.
FASTON terminal
connector
To accomplish installation, reverse the removal procedure.
009-058-00
10.20.2000 FM2732 (2)
Cam
WP3x6 (Loktite)
Motor shaft
(flat-machined surface)
Ring
FR7H36AQ.EPS
CR-IR347 Service Manual
Motor
Marking
FR7H36AO.EPS
MC - 337
MC - 337