Active Alarm
No Active Alarms
2000966-386D
The number of times the Silence Alarm key is pressed.
Whether any alarms are active when the Silence Alarm key is pressed.
Alarms remained paused unless a CRISIS patient status alarm occurs. Alarm pause
breakthrough allows any CRISIS patient status alarm to interrupt an alarm pause.
Only CRISIS alarms breakthrough, other patient status alarms will not sound.
When the monitor is in alarm pause:
Countdown timer — Displays time left in the alarm pause when in ADULT-ICU
or NEONATAL-ICU patient-monitor types.
NBP alarms — Silenced until the next measurement is taken.
ECG or arrhythmia alarms — If a patient condition of equal or greater priority
occurs, the alarm pause is cancelled and the alarm will "break through".
When ECG is turned off, the "ALARM PAUSE" system status message does not
display on the central station.
FRANCE locale — Alarms are paused for no more than three minutes in
ADULT-ICU and NEONATAL-ICU patient-monitor types.
Key Pressed
One
60 seconds
Two
5 minutes
Three
Pause cancelled
Four
Five
NOTE
You can also select Silence Only from the monitor defaults. This disables the
pause and a second push results in reactivation of alarm (versus an extended
pause).
Key Pressed
One
5 minutes
Two
Pause cancelled
Three
Four
Dash™ 3000/4000/5000
ADULT-ICU
NEONATAL-ICU
3 minutes
ADULT-ICU
NEONATAL-ICU
3 minutes
Alarms
OPERATING
ROOM
5 minutes
15 minutes
Permanent pause
Pause cancelled
OPERATING
ROOM
5 minutes
15 minutes
Permanent pause
Pause cancelled
4-7