Monitoring ECG
8-10
Option
DISPLAY
ECG SIZE
DETECT PACE
ECG LIMITS
VIEW ALL ECG
RELEARN
ST ANALYSIS
12 LEAD ECG ANALYSIS
ECG FILTER
Dash™ 3000/4000/5000
Function
Select the primary (top trace) lead. The V lead choice is
determined by the V lead identified in the ST menu.
Select the waveform size.
1X — Recommended for standard monitoring.
2X, 4X — Recommended for low amplitude QRS
waveforms. However, this lowers the QRS detection
threshold and baseline artifact may be detected as a
QRS.
Turn pacemaker detection on or off.
Select the alarm limits for heart rate and PVCs.
NOTE
PVC Limit is available in the CARDIAC software
package only.
Displays all six leads of ECG.
Relearn the patient's ECG pattern to correct arrhythmia calls
and heart rate value, restore ST measurements and enable
more accurate monitoring.
Turn the ST analysis program on or off. For more
information, refer to
Performing ST Analysis on page
Displays 12 leads of ECG. For more information, refer to
Performing 12 Lead Analysis on page
Select a ECG waveform signal filter to improve the
waveform display and graphs. (Does not affect ST and
arrhythmia analysis.)
DIAGNOSTIC — Provides the maximum available
ECG signal information. Recommended for high-fidelity
recordings in low-noise environments.
MONITORING — Provides modest reduction of high
frequency interference. Recommended for typical
monitoring applications.
MODERATE — Provides significant reduction of high
frequency interference. Recommended for high-noise
environments (for example, electrosurgical units).
MAXIMUM — Provides significant reduction of high
frequency interference plus maximum stabilization of
the ECG baseline. This mode alters the displayed QRS
morphology, but does not affect automated analysis.
Do not rely upon the displayed QRS morphology for
diagnostic purposes when this mode is selected.
WARNING
—The MAXIMUM filter alters displayed ECG morphology.
Do not rely on ECG morphology for diagnostic purposes
when this filter is selected.
NOTE
MODERATE and MAXIMUM filters are not
recommended for pacemaker patients.