ECG Setting Source When Entering COMBO Mode
Monitor
Not admitted
Admitted
Not admitted
Monitor
Not admitted
Admitted
Not admitted
2000966-386D
Arrhythmia alarm levels
HR, PVC and ST parameter alarm levels
ECG SIZE
DETECT PACE
ARRHYTHMIA
ST ANALYSIS
LEAD ANALYSIS
HR HIGH LIMIT and HR LOW LIMIT
PVC LIMIT and PVC state
To select the source of ECG settings, follow this procedure.
1.
Select MORE MENUS > MONITOR SETUP > MONITOR DEFAULTS >
SETUP DEFAULT DISPLAY.
2.
Locate and select COMBO DEFAULT SOURCE.
3.
Select one of the following:
TELE DEFINED — Use the telemetry ECG settings. This is the factory
default.
USER DEFINED — Use the monitor's ECG settings if telemetry is NOT
admitted. Use the telemetry ECG settings if telemetry IS admitted.
The following tables indicate the source (monitor or telemetry) of ECG settings when
entering COMBO mode.
Tele Defined
Telemetry
Not admitted
Not admitted
Admitted
User Defined
Telemetry
Not admitted
Not admitted
Admitted
Notice that selecting Tele Defined means the ECG setting source is always telemetry
when entering COMBO mode, while the ECG setting source for User Defined
depends on whether the patient is admitted to the monitor or telemetry when entering
COMBO mode.
Dash™ 3000/4000/5000
Entering COMBO Mode
>
>
>
Entering COMBO Mode
>
>
>
Monitoring ECG
ECG Setting Source
Telemetry
ECG Setting Source
Monitor
Telemetry
8-13