Rf Output Level Linearity Test (@ Port1) - Agilent Technologies E5071C Service Manual

Ena rf network analyzers
Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

Frequency [Hz]
Test limit [dB]
8.025 G
8.5 G

RF Output Level Linearity Test (@ Port1)

1. Power Sensor: E9304A Option H18
CW Frequency: 9 kHz (relative to 0 dBm reference)
Option 230, 240, 280, 430, 440, 480
Power level
Test limit [dB]
[dBm]
-20
-17.5
-15
-12.5
-10
-7.5
-5
-2.5
2.5
5
7.5
10
CW Frequency: 100 kHz (relative to 0 dBm reference)
Power level
Test limit [dB]
[dBm]
-20
-17.5
-15
-12.5
-10
Chapter 2
E5071C Performance Test Record
Test result
± 1.0
± 1.0
Test result [dB]
± 0.75
± 0.75
± 0.75
± 0.75
± 0.75
± 0.75
± 0.75
± 0.75
± 0.75
± 0.75
± 0.75
± 0.75
Test result [dB]
± 0.75
± 0.75
± 0.75
± 0.75
± 0.75
Performance Test
[dB]
Measurement
uncertainty [dB]
± 0.30
± 0.30
Measurement
uncertainty [dB]
± 0.23
± 0.23
± 0.23
± 0.23
± 0.23
± 0.22
± 0.22
± 0.22
± 0.21
± 0.21
± 0.21
± 0.21
Measurement
uncertainty [dB]
± 0.23
± 0.23
± 0.23
± 0.23
± 0.23
65