Agilent Technologies E5071C Service Manual page 162

Ena rf network analyzers
Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

Performance Test
E5071C Performance Test Record
System
Frequency
performance
range [Hz]
Source match
3 G - 6 G
Source match
6 G - 8.5 G
Load match
9 k - 300 k
Load match
300 k - 3 G
Load match
3 G - 6 G
Load match
6 G - 8.5 G
Transmission
9 k - 300 k
Tracking
Transmission
300 k - 3 G
Tracking
Transmission
3 G - 6 G
Tracking
Transmission
6 G- 8.5 G
Tracking
Reflection
9 k - 300 k
Tracking
Reflection
300 k - 3 G
Tracking
Reflection
3 G - 6 G
Tracking
Reflection
6 G- 8.5 G
Tracking
Port 4
System
Frequency
performance
range [Hz]
Directivity
9 k - 300 k
Directivity
300 k -3 G
Directivity
3 G - 6 G
Directivity
6 G - 8.5 G
Source match
9 k - 300 k
Source match
300 k -3 G
Source match
3 G - 6 G
Source match
6 G - 8.5 G
162
Test limit
Test result
[dB]
[dB]
< -20
< -15
< -12
< -17
< -12
< -10
± 1.5
± 1.0
± 1.0
± 1.0
± 1.5
± 1.0
± 1.0
± 1.0
Test limit
Test result
[dB]
[dB]
< -20
< -25
< -20
< -15
< -20
< -25
< -20
< -15
Measurement
uncertainty [dB]
1.94
0.57
0.04
0.59
0.66
0.64
± 0.03
± 0.20
± 0.24
± 0.35
± 0.04
± 0.04
± 0.66
± 0.64
Measurement
uncertainty [dB]
0.09
0.81
0.40
0.67
0.05
3.00
1.94
0.57
Chapter 2